Gas Scavenging Inlet, Venturi Type (AGSS)
Brand - Silbermann, LLC
The Silbermann Gas Scavenging Inlet venturi-type is designed to safely remove surplus anesthetic gas and explosive gas mixtures from the operating theater.
The Silbermann jet scavenging inlet is designed to meet the latest and most stringent NFPA 99 requirements.
The scavenging inlet assembly uses compressed air, which creates suction (Venturi) and propels the contaminated air to the outside atmosphere.
The jet scavenging inlet can be supplied recessed, panel-mounted, or exposed, with or without a hose assembly.
To connect equipment, you can select and purchase compatible adapters and plugs.
Operation based on compressed air – No need for a central AGSS system.
When the probe is connected, a green pneumatic indicator showing the system is in use.
Very quiet operation.
Automatic shut off when the probe is disconnected.
Flow and vacuum adjustment.
The locking mechanism quick connect.
Medical Gas System Standards: NFPA99, ISO 7396, G-01, HTM 02-01.
Silbermann’s Quality Management System: EN ISO 13485-2016, ISO 9001-2015, CE marked.
inlet pressure: 3.5-5 bars
Inlet flow: 40 LPM
Free air suction capacity: Adjustable 35 -80 LPM
Vacuum level: Adjustable 20-50 cmH2O
Inlet pipe diameter: OD 3/8”
Exhaust pipe diameter: OD 7/8”